Connection of Resistance Thermometers in 3-wire Configuration

When resistance thermometers (Pt100, Pt1000, Ni1000) are used, a constant current must flow through them to build the necessary voltage drop for the evaluation. For this, the module CI511-ETHCAT provides a constant current source which is multiplexed over the max. 4 analog input channels.

The following figure shows the connection of resistance thermometers in 3-wire configuration.

../_images/81c3504d6b04038c0a33139059c84da1

Connection of resistance thermometers in 3-wire configuration

With 3-wire configuration, two adjacent analog channels belong together (e. g. the channels 0 and 1). In this case, both channels are configured according to the desired operating mode. The lower address must be the even address (channel 0), the next higher address must be the odd address (channel 1).

The constant current of one channel flows through the resistance thermometer. The constant current of the other channel flows through one of the cores. The module calculates the measured value from the two voltage drops and stores it under the input with the higher channel number (e. g. I1).

In order to keep measuring errors as small as possible, it is necessary, to have all the involved conductors in the same cable. All the conductors must have the same cross section.

Pt100 -50 °C…+400 °C 3-wire configuration, 2 channels used
Pt1000 -50 °C…+400 °C 3-wire configuration, 2 channels used
Ni1000 -50 °C…+150 °C 3-wire configuration, 2 channels used

The measuring ranges are described in the section Measuring Ranges CI511-ETHCAT CI511-ETHCAT.

The module CI511-ETHCAT performs a linearization of the resistance characteristic.

In order to avoid error messages from unused analog input channels, it is useful to configure them as “unused”.